Accession Number ADA584277
Title Use of Field Deployable Instrumentation for the Monitoring of Munitions Constituents in Groundwater.
Publication Date Mar 2011
Media Count 20p
Personal Author A. Bednar A. Russell C. Hayes D. Splichal T. Georgian
Abstract Field portable instrumentation can provide near real time analysis of munitions constituents in water. Quantitative Agreement for Most Analytes. TNB was positively biased possibly due to elevated TNT. RDX is difficult by thermal separation methods. Limited to screening level data without further refinement. Method development and instrument optimization are critical.
Keywords Agreements
Ammunition
Briefing charts
Deployment
Field equipment
Ground water
Instrumentation
Mass spectrometers
Monitoring
Portable equipment
Rdx
Real time
Refining
Separation
Thermal properties
Time series analysis
Tnt
Water


 
Source Agency Non Paid ADAS
NTIS Subject Category 68D - Water Pollution & Control
79A - Ammunition, Explosives, & Pyrotechnics
Corporate Author Army Engineer Waterways Experiment Station, Vicksburg, MS. Environmental Lab.
Document Type Technical report
Title Note Briefing charts.
NTIS Issue Number 1402
Contract Number N/A

Science and Technology Highlights

See a sampling of the latest scientific, technical and engineering information from NTIS in the NTIS Technical Reports Newsletter

Acrobat Reader Mobile    Acrobat Reader